Bronkhorst® model F-201DV Mass Flow Controllers (MFCs) are suited for precise measurement of flow ranges between 0,42…21 mln/min and 0,042…2,1 ln/min (N2-equivalent). The instruments are particularly suited for corrosive gases or applications with very low differential pressure (∆P). Compared to conventional instruments, LOW-∆P-FLOW MFCs have larger flow channels to minimize the risk of clogging, facilitate cleaning and purging, and cause lower pressure drop (the sensor only requires 0,5 to 5 mbar).

The integrated digital pc-board provides signal and fieldbus conversion as well as PID controller functionality for mass flow control by means of the integrated control valve. In addition to the standard RS232 output the instruments also offer analog I/O. As an option, an on-board interface can be mounted to provide CANopen®, DeviceNet™, EtherCAT®, PROFIBUS DP, PROFINET, Modbus RTU, ASCII or TCP/IP, EtherNet/IP, POWERLINK or FLOW-BUS protocols.


Technical specifications

Measurement / control system
Flow range (intermediate ranges available)min. 0,42…21 mln/min
max. 0,042…2,1 ln/min
(based on N2)
Accuracy (incl. linearity) (based on actual calibration)± 1 % FS
Repeatability< 0,2 % RD
Turndown ratio1:50 (2…100%)
Max. operating pressure10 bar g
Multi fluid capabilityStorage of max. 8 calibration curves
Settling time (in control, typical)2 … 3 sec.
Control stability< ± 0,1 % FS (typical)
Operating temperature-10 … +70 °C
Mountinghorizontal
Temperature sensitivity< 0,1% FS/°C
Pressure sensitivity0,1% Rd/bar typical N2
Max. Kv-value6,6 x 10-2
Leak integrity, outboardtested < 2 x 10-9 mbar l/s He
Warm-up time30 min. for optimum accuracy
2 min for accuracy ± 2% FS
Mechanical parts
Material (wetted parts)stainless steel 316L or comparable;
other on request
Process connectionscompression type or face seal (VCR/VCO) couplings
Sealsstandard: FKM/Viton®;
options: EPDM, FFKM/Kalrez®, FDA and USP Class VI approved compounds
Weight0,6 kg
Ingress protectionIP40
Electrical properties
Power supply+15 … 24 Vdc
Max. power consumption
Supply
15 V
24 V
at voltage I/O
290 mA
200 mA
at current I/O
320 mA
215 mA
extra for fieldbus
<75 mA
<50 mA
Analog output0…5 (10) Vdc or 0 (4)…20 mA (sourcing output)
Digital communicationstandard: RS232;
options: PROFIBUS DP, CANopen®, DeviceNet™, PROFINET, EtherCAT®, Modbus RTU, ASCII or TCP/IP, EtherNet/IP, POWERLINK, FLOW-BUS
Electrical connection
Analog/RS2329-pin D-connector (male);
PROFIBUS DPbus: 9-pin D-connector (female);
power: 9-pin D-connector (male);
CANopen® / DeviceNet™5-pin M12-connector (male);
FLOW-BUS/Modbus-RTU/ASCIIRJ45 modular jack
Modbus TCP / EtherNet/IP / POWERLINK2 x RJ45 modular jack (in/out);
EtherCAT®/ PROFINET2 x RJ45 modular jack (in/out);
Control valve options
External actuator options to be connected to the controller
Certification for hazardous areas
Approvals / certificates

Technical specifications subject to change without notice.

For dimensional drawings and hook-up diagrams please visit the product page on our website

Downloads

Brochures
LOW-ΔP-FLOW Brochure
Manuals
General manual digital instruments
Operating instructions digital instruments
Quick start guide LOW-dP-FLOW industrial style
Hook-up diagrams
Analog IO - RS232
DeviceNet
EtherCAT
FLOW-BUS
Modbus RTU/ASCII
PROFIBUS DP
PROFINET
CANopen
EtherNet/IP
Modbus-TCP
POWERLINK
Optional Bus and IO Configurations
Dimensional drawing
Dimensional drawing F-201DV_F-201EV
Solids 3D
F-201DV


台灣柏朗豪斯特股份有限公司

新竹縣竹北市台元街三十八號四樓之一、二
Rm. 1 & 2, 4F., No.38, Taiyuan St., Zhubei City, Hsinchu County 302, Taiwan (R.O.C.)
電話 +886 3-5600-518
[email protected]

            Social Responsibility
Copyright © 2024 Bronkhorst. All rights reserved.     Sitemap     Disclaimer     Privacy note