Bronkhorst

EX-FLOW F-231MX

氣體用高壓質量流量計,防爆型

  • ATEX 認證 Cat.2,Zone 1
  • 精度高,重複性好
  • 幾乎與壓力和溫度無關
  • 工作壓力 / delta-P 高達 350 bar
詢價 下載 數據表
EX-FLOWF-231MX

Intrinsically Safe Gas Mass Flow Controllers for High Pressure / High Delta-P

Bronkhorst® EX-FLOW Mass Flow Controllers (MFCs) are suited for precise gas flow control in ATEX Zone 1 hazardous areas. The flow meter part and the control valve of the MFC should be connected (via separate cables) to a power supply with galvanic isolation / preamplifier / readout system (located in the safe zone) which contains a controller board to complete the control loop. See Bronkhorst® E-8000 Series.

EX-FLOW model F-231MX covers flow ranges from 10...500 mln/min up to 0,2...10 ln/min (N2-equivalent) at operating pressures up to 350 bar and max. 350 bar pressure difference (ΔP).

The intrinsically safe measuring head of the flow meter is tested according to ATEX 114 Directive 2014/34/EU and approved under EC-Type Examination Number: KEMA 01ATEX1172, protection II 2 G Ex ib IIC T4 Gb.
The intrinsically safe valve coils are explosion proof certified and available in two options (ATEX certification only):
Coil XB: protection II 1 G Ex ia IIC T6 / protection II 1 D Ex ta IIIC T80°C
Coil XC: protection II 2 G Ex eb IIC T4 / protection II 2 D Ex tb IIIC T130°C


Technical specifications

Measurement / control system
Flow range (intermediate ranges available)min. 10…500 mln/min
max. 0,2…10 ln/min
(based on N2)
Accuracy (incl. linearity) (based on actual calibration)± 1 % FS
Repeatability< 0,2 % RD
Turndown ratio1:50 (2…100%)
Time constant5 sec.
Control stability≤ ± 0,1 % FS typical
Operating temperature-10 … +65 °C
Temperature sensitivityzero: < 0,05% FS/°C; span: < 0,05% Rd/°C
Leak integrity, outboardtested < 2 x 10-9 mbar l/s He
Attitude sensitivitymax. error at 90° off horizontal 0,2% FS at 1 bar, typical N2
Warm-up time30 min. for optimum accuracy
2 min. for accuracy ± 2% FS
Mechanical parts
Material (wetted parts)Stainless steel 316L or comparable
Pressure rating (PN)350 bar abs
Min. ΔP6 bar dif.
Max. ΔPup to 350 bar dif.
Process connectionscompression type or face seal (VCR/VCO) couplings
Sealsstandard: FKM/Viton®;
options: EPDM, FFKM/Kalrez®
Ingress protectionIP65
Electrical properties
Output signal15 … 20 mA (linear)
Terminal connection, cable gland M16x1,5
I/O signals via PS/Readout (located in safe area)analog: 0…5 Vdc, 0…10 Vdc, 0…20 mA, 4…20 mA;
digital: RS232, PROFIBUS DP, DeviceNet™, Modbus RTU or ASCII, PROFINET, EtherCAT®, FLOW-BUS
XB-coilCoil voltage max. 28 V/110 mA;
295 Ohm at 20°C
XC-coilCoil voltage max. 24 V;
65 Ohm at 20°C, Pmax = 9 W at 20°C
Electrical connection
Ex-proof measuring headTerminal connection, cable gland M16x1,5
Control valve coilcable gland M20x1,5
Control valve options
External actuator options to be connected to the controller
Certification for hazardous areas
Approvals / certificates

Technical specifications subject to change without notice.

For dimensional drawings and hook-up diagrams please visit the product page on our website

Downloads

Brochures
EX-FLOW brochure
Manuals
Manual EX-FLOW meters and controllers - ATEX
Manual EX-FLOW meters and controllers - IECEx
Manual EX-FLOW meters and controllers - TIIS
Hook-up diagram
MFM Measuring Head
XB-Coil
XC-Coil
Dimensional drawing
Dimensional drawing F-230MX-F-231MX-F-232MX
Certificates
IECEx Certificate EX-FLOW MFM Measuring Head X100 : IECEx DEK14.0060
TIIS Certificate EX-FLOW MFM Measuring head X100 : 検・第TC21584号 (Japanese certification)
KCs Certificate EX-FLOW MFM Measuring Head X100 :: Ex ib IIC T4 (Korean certification)
ATEX CAT 2 Certificate EX-FLOW MFM Measuring Head X100 : KEMA 01ATEX1172
EU Declaration of Conformity Measuring Head X100 : ATEX CAT 2 (doc. 9.06.032)
ATEX CAT 2 Certificate XB-coil
ATEX CAT 2 Certificate XC-coil
ATEX CAT 2 Notification Directive 2014/34/EU : DEKRA 12ATEXQ0100


台灣柏朗豪斯特股份有限公司

新竹縣竹北市台元街三十八號四樓之一、二
Rm. 1 & 2, 4F., No.38, Taiyuan St., Zhubei City, Hsinchu County 302, Taiwan (R.O.C.)
電話 +886 3-5600-518
[email protected]

            Social Responsibility
Copyright © 2024 Bronkhorst. All rights reserved.     Sitemap     Disclaimer     Privacy note