Bronkhorst

mini CORI-FLOW Ex d XM12

방폭 코리올리스 질량 유량계 / 컨트롤러

  • 높은 정확도, 높은 재연성
  • 빠른 반응
  • 유체 물성에 독립적임
  • 밀도 / 온도 출력 추가 옵션
  • IECEx 및 ATEX Zone 1 승인
Get a quote 다운로드 Datasheet
mini CORI-FLOW Ex dXM12

Ex-proof Coriolis Mass Flow Meters/Controllers for low flow rates

mini CORI-FLOW Ex d Mass Flow Meters are precise and compact instruments for low flow rates, based on the Coriolis measuring principle, built in an explosion proof housing for use in IECEx and ATEX Zone 1 hazardous areas, ATEX approval II 2 G Ex d e IIB T6 Gb. Bronkhorst® model XM12 Mass Flow Meter (MFM) is suited for highly accurate measurement of gas or liquid flow in the range of 0…200 g/h (which corresponds with 0…2,66 ln/min when used on nitrogen) at operating pressures up to 138 bar (a).

The instrument contains a microprocessor based pc-board with signal and fieldbus conversion and a PID controller for optional mass flow control by means of a separately mounted control valve or pump.


기술 정보

Measurement / control system
유량Liquid: 0…200 g/h (nominal flow rate: 100 g/h);
Gas: 0…2,66 ln/min (N2);
Full Scale (FS) value is user-configurable
Mass flow accuracyLiquid: ±0,2% Rd;
Gas: ±0,5% Rd
Repeatability±0,05% of rate ± ½(ZS* x 100/actual flow)%
Turndown ratioMeter: up to 1:100;
Controller: ≥ 1:50
Zero stability (ZS)< ± 0,02 g/h
(Guaranteed at constant temperature and for unchanging process and environmental conditions.)
Response time (sensor)200 msec
Settling time (in control, typical)1 sec. (typical)
Temperature rangeAmbient temperature range: 0…55°C;
Process temperature range: 0…70°C
Temperature effecton zero: < 0,01 g/h/°C;
on span: < 0,001% Rd/°C;
self heating (at zero flow): < 15°C
(Depends on flow rate, heat capacity fluid, T amb., T fluid and cooling capacity.)
MountingAny position, attitude sensitivity negligible.
External shocks or vibrations should be avoided.
Temperature accuracy± 0,5°C
Density accuracy< ± 5 kg/m³ (at full scale flow)
Leak integrity, outboardtested < 2 x 10-9 mbar l/s He
Warm-up time> 30 min. for optimum accuracy
Mechanical parts
Sensorsingle tube, DN 0,25
Valve seatKalrez®-6375, other on request
(for controllers)
Material (wetted parts)stainless steel 316L or comparable
Pressure rating (PN)138 bar a
Process connections1/8" OD compression type (welded);
other on request
Sealsmetal
Ingress protectionIP66
Electrical properties
Power supply+15…24 Vdc +/- 10%
Max. ripple recommended: 50 mV tt
Max. power consumptionMeter: max. 3 W;
Controller: max. 7 W
아날로그 출력0…5 (10) Vdc, min. load impedance > 2 kΩ;
0 (4)…20 mA (sourcing), max. load impedance < 375 Ω;
on request: Ex i output 4…20 mA
아날로그 설정값(for MFM + pump or control valve)
0…5 (10) Vdc, min. load impedance > 100 kΩ;
0 (4)…20 mA (sourcing), max. load impedance ~ 250 Ω
Digital communicationstandard: RS232
options: PROFIBUS DP, DeviceNet™, Modbus RTU or FLOW-BUS
Electrical connectionAll instrument connections are wired to screw terminals
Electrical connection
Control valve options
External actuator options to be connected to the controller
Electromagnetic control valve (Bronkhorst®)C2I valve with XC coil or XB coil through barrier
Pneumatic actuated control valve (Badger Meter)RC200 valve with Ex d TEIP11 I/P converter
Mass Flow Meter controlled pumpPump with ATEX zone 1 U/f converter
Certification for hazardous areas
Approvals (MFM)ATEX: II 2 G Ex d e IIB T6 Gb | DEKRA 12ATEX0144X |
IECx: Ex d e IIB T6 Gb | IECEx DEK 12.00400 |
TIIS Zone 1 / JP Ex IIB T6 Gb | DEK 19.0041 |
Approvals / certificates

Technical specifications subject to change without notice.

도면 및 배선연결 정보 제품 페이지 우리의 웹사이트

Downloads

Brochures
mini CORI-FLOW Ex d brochure
Manual (incl. dimensions)
mini CORI-FLOW Ex d - Manual
Certificates
ATEX: DEKRA 12ATEX0114X
IECEx: IECEx DEK 12.0040
TIIS Zone 1 / JP Ex: DEK 19.0041


BRONKHORST KOREA Co. Ltd

경기도 화성시 동탄순환대로 830 SK
V1 Center 209호
대표자 : 헨드리크제지너스타펠
사업자등록번호 : 119-86-74469

           Social Responsibility